YL9300 HPLC system
Excellent flow precision < 0,1%
Wide range of flow from 1 ul / min up to 5 ml/min ( analytical HPLC ) and from 1 ul /min up to 10 ml/min ( option)
Automatic piston back flushing
Quaternary solvent system including on line degasser
Automated Solvent compressibility compensation
Easy maintenance and Auto diagnostic at start up
Multi parameter monitoring possibility and much more...
Included UV-VIS detector:
Wavelength range: 190-600 nm (Standard) 190-900nm (Option)
Data collection rate: up to 50Hz
Light source: Deuterium arc lamp / Tungsten lamp addible for visible range (Option)
2-Order filter: Automatic filter switching
Bandwidth: 5 nm
Wavelength accuracy: ±1 nm, Wavelength precision: ±0.1nm, Linearity: >99.5% at 2.5 AU (acetone, 254nm)
Noise level: <±0.35 X 10-5 Abs/min, 254nm, dry cell ,Drift : <1 X 10-4 Abs/hour
Dispersion element: Concave grating, Seya-Namioka
Flow cell design: Cone type,Path length: 10 mm (Analytical cell) volume: 10 uL
Pressure limit: 2000 psi (Analytical cell )
Optics and electronic diagnostic routine
Wavelength Verification: On demand via internal holmium filter
Analytical HPLC all in one
YL9110 Quaternary Pump
Operating principle: Parallel dual-plunger pump, Low-pressure gradient
Control mode: Constant flow, Constant pressure
Compressibility compensation: Automatic
Flow range: 0.001-10ml/min
Flow rate accuracy : ≤1% at 1ml/min
Flow rate precision: 0.1% RSD at 1ml/min
Maximum pressure: 6000 psi > Operating range: 0-6000 psi up to 5ml/min > Operating range : 0-3000 psi up to 10ml/min
Pressure pulsation : ≤1% at 1ml/min
Number of eluent lines: 4
Gradient formation: 4-channel mixing valve
Composition Precision: <0.1%
Composition Accuracy: <0.5%
Auto prime/purge
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, automatic voltage selection, 50/60Hz, ±5% , 70W
YL9111 Binary Pump
Operating principle: X 2 Parallel dual-plunger pump, High-pressure gradient
Flow range: 0.001-10ml/min
Number of eluent lines: 2
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 140W
YL9120 UV/Vis Detector
Wavelength Range: 190-900 nm
Data collection rate: up to 50Hz
Deuterium arc lamp & tungsten lamp
2-Order filter: Automatic filter switching
Bandwidth: 5.5 nm
Wavelength Accuracy : ±1 nm
Wavelength Precision : ±0.1nm
Linearity: >99.5% at 2.5 AU (acetone, 254nm)
Noise level : <±0.5 X 10^(-5) Abs/min , 254nm, dry cell
Drift : <1 X 10^(-4) Abs/hour
Dispersion element: Concave grating
Warm-up period: 1 hour
Flow Cell Design: Cone type
Path Length: 10 mm (Analytical cell)
Cell Volume: 10 uL (Analytical cell)
Pressure limit: 1500psi (Analytical cell)
Wetted materials: 316 stainless steel, Quartz, Teflon
Analog Outputs: Two, software selectable: Absorbance Energy, Sample Energy, Reference Energy Communications: LAN
Power-up Diagnostics: Optics and electronic diagnostic routine
Power-up Wavelength Verification: Automatic on power up via internal holmium filter and D2 lamp
5-Point Wavelength Calibration: On demand via internal holmium filter and D2 lamp
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 100W
YL9160 PDA Detector
PDA A/D Resolution: 16bit
Slit Bandwidth: 1.7nm Pixel Resolution: 0.9nm Channel: 1024
Wavelength: 190~950
Shield Optics
Analytical Cell
- Path-length: 10mm
- Pressure : >1500psi
- Volume : 13ul
Dimension: W: 385mm H: 160mm D: 565mm
Weight: 15Kg
Noise Level: +/- 0.00002 AU (Empty Cell, 1sec Rise Time, 254nm)
Drift : 0.0002AU/hr (Baseline Correction), 0.001AU/hr (Room Temp)
Wavelength Accuracy: <1nm (HY-1 Holmium Oxide Filter)
Sampling Rate: Max. 50Hz
1~16 nm Spectral Bandwidth, Wide Spectral Range, Alternative Spectral Function
GLP Compliance:
- Photometric Accuracy, Linearity, Noise Level, Drift - System Check,
Analog Output: 2 channel (Replaceable Dual Wavelength Detector)
Valve Output: 2ch, Programmable Sampling (Fraction Collector Function Available )
Trigger Input: 2ch
Communication: TCP/IP, RS232C
Simultaneous Data Channel: 8 or Full Scan
Filtering: Bessel, RC, Aver. 0.01~10Hz
YL9150 Autosampler (Alias)
Sound pressure level: LeAq < 70 dB
Working temperature: 10 - 40°C
Humidity: 20 - 80% RH
Safety and EMC compatibility: According to EC-directives; CSA (UL) approved
Power requirements: 95 - 240 Volt AC ± 10%; 50/60 Hz; 200VA
Sampling:
Sample capacity: 2 Micro Well Plates according to SBS standards; 96-well high/low
Loop volume: 1 - 5000 µL programmable, 10 mL loop optional
Dispenser syringe: 500 µL standard or 2500 µL for Prep option
Vial detection: Missing vial/well plate detection by sensor
Headspace pressure: Built-in compressor, but only for vials with septa
Switching time injection valve: Electrically < 100 msec
Piercing precision needle: ± 0.6mm
Wash solvent: External wash solvent bottle (bracket optional)
Wetted parts in flow path: SS316, PTFE (teflon), ETFE (tefzel), VESPEL, PCTFE (Kel-F). Optional: PEEK
(10) Injection cycle time: < 60 sec. in all injection modes for 1 injection ≤ 100 µL including 300 µL wash / < 20 sec in partial loopfill mode for 1 injection of 10µL including 500 µL wash
YL9101 Vacuum Degasser
Number of channel: 4 CHs
Maximum flow rate: 10ml/min per channel
> 0-2.0ml/min per channel for 70% Gas Removed from Methanol
Internal volume per channel: 925ul per channel
Materials in contact with solvent: Teflon AF, PEEK and Glass-filled PTFE
Safety & maintenance: Error detection
Dimensions: 385 X 80 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, automatic voltage selection; 50/60Hz, ±5%, 20W
Download info